Quality Assurance Program Assistance and Audits
AC Customers
|
QA Program AssistanceThings To KnowKnow what regulations apply to you and follow the regulations. Every state and country is different.
Have a commensurate with scope QA plan and follow it along with everything else the overseeing entity gave you to do. 1 set of passive devices (duplicates) counts as 2 devices. So every 50 sets of dups would require you to do spikes. A Spike chart has a control limit of 30% but somehow some folks are using 25% - which is more restrictive anyway. Use door signs and get closed house agreements signed. Communicate with all parties involved. Explain the process and what their responsibility is. Closed house conditions SHALL be in place 12 hours before you start a test lasting less than 4 days. Pay attention to CHC and test appropriately. If your equipment has a 12 hour delay you could establish CHC, set device to delay and then run just a 2 day test. You may have to close up the house and come back in 4 days instead of 2 days so communicate clearly so your scheduling does not get messed up. Use the free resources. Network and educate don't tolerate. Remember that you are the expert so make sure that people let you do your job-you are the one that will get in trouble. Keep your exposure ALARA. |